Shimadzu Recalls Trinias Interventional X-Ray Systems Due to Loose Mounting Bolts
Shimadzu Corporation is recalling 118 X-ray systems because C-arm drive unit mounting bolts may not be torqued to specification.
Shimadzu Corporation Medical Systems Division has recalled 118 units of its Trinias Opera Smart, Opera Grand, and Smart OBL Interventional Fluoroscopic X-Ray Systems.
The recall was initiated because the mounting bolts securing the C-arm drive unit may not be torqued to specification. This issue can compromise the structural stability and safe operation of the medical imaging system.
The recalled devices were distributed to California, Florida, New York, South Carolina, Tennessee, and Texas. The U.S. Food and Drug Administration has classified this action as a Class II recall.
What to do
- Stop using the affected interventional fluoroscopic X-ray systems immediately.
- Contact Shimadzu Corporation Medical Systems Division for guidance on inspecting and servicing the equipment.
- Report any adverse events or quality problems regarding the device to the FDA.
At a glance
| Agency | U.S. Food and Drug Administration (FDA) |
|---|---|
| Recall number | Z-0347-2027 |
| Date | 2026-11-11 |
| Company | SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION, Kyoto |
| Severity | Class II |
| Quantity | 118 unts. |
| Where sold | US Nationwide distribution in the states of SC, TX, FL, TN, CA, FL, NY. |
| Lot codes / identifiers | Serial number/Shimadzu Medical Systems #/UDI: M1F0D3BF4001/ SMS20250167/ (01)04540217067145(11)250; M1F0CC4EB001/ SMS20250098/ (01)04540217067022(11)241; M1F0D27EA001/ SMS20250001/ (01)04540217067138(11)241; 41F0F43E7001/ SMS20250034/ (01)04540217068951(11)240; M1F0D3BE7001/ SMS20250248/ (01)04540217067145(11)240; M1F0D3BE6001/ SMS20240196/ (01)04540217067145(11)240; 41F0F40E5001/ N/A/ (01)04540217068937(11)240; 41F0F41E5001/ N/A/ (01)04540217068944(11)240; 41F0F43E5001/ SMS20240236/ (01)04540217068951(11)240; M1F0CD7E4001/ SMS20240162/ (01)04540217067046(11)240; 41F0F41E1001/ SMS20240062/ (01)04540217068944(11)240; 41F0F43DB002/ SMS20240101/ (01)04540217068951(11)231; 41F0F43DB001/ SMS20240100/ (01)04540217068951(11)231; M1F0CD7D4001/ SMS20230178/ (01)04540217067046(11)230 |
Source: official FDA notice. This summary was written by 365 News from the official notice. Details can change; check the notice before acting.