BioMerieux Recalls EMAG Systems Over Software Error Causing Delayed Test Results
BioMerieux SA is recalling certain EMAG systems due to a software error that can cause delayed or missing clinical sample results.
BioMerieux SA has recalled 33 EMAG system automated nucleic acid sample preparation instruments (Model/Catalog Number 418591/418591U).
The recall was initiated due to a randomly occurring software error, designated as Error 10003 Heater timeout, which appears after updating to firmware SP4 Patch 2. This error can result in delayed or missing test results when clinical samples are being analyzed.
The instruments were distributed worldwide, including U.S. distribution in California, Indiana, Kansas, Louisiana, Minnesota, Missouri, New York, Ohio, and Texas. Distribution also included Canada, the United Kingdom, Sweden, France, Norway, the Netherlands, Belgium, Denmark, Germany, and Austria. The recall is classified as a Class II recall by the FDA.
What to do
- Stop using affected EMAG systems if they are experiencing software errors or delayed sample results.
- Contact BioMerieux SA for further guidance and technical support regarding the firmware update and software error.
- Report any issues or adverse events related to the instrument to the FDA.
At a glance
| Agency | U.S. Food and Drug Administration (FDA) |
|---|---|
| Recall number | Z-0301-2027 |
| Date | 2026-11-04 |
| Company | BioMerieux SA, Marcy L’Etoile |
| Severity | Class II |
| Quantity | 33 |
| Where sold | Worldwide – US Nationwide distribution in the states of CA, IN, KS, LA, MN, MO, NY, OH, and TX. The countries of Canada, United Kingdom, Sweden, France, Norway, Netherlands, Belgium, Denmark, Germany, Austria. |
| Lot codes / identifiers | UDI-DI 357302BUDI000544TP Serial Numbers IM03343 IM03283 IM03104 IM03329 IM03376 IM03474 IM03321 IM03071 IM03175 IM03194 IM03515 IM03516 IM03240 IM03478 IM03479 IM03482 IM03034 IM03092 IM03378 IM03414 IM03177 IM03193 IM03506 IM3054 IM03439 IM03350 IM03380 IM03131 IM03148 IM03153 IM03275 IM03314 IM03351 |